PRESS RELEASE

Windstone Medical Packaging dba Aligned Medical Solutions Issues Nationwide Recall of Pharmacy Lidocaine HCL 1% 5mL SDV placed into the following kit: AMS16835

 

MAY 7, 2026

1602 4TH AVENUE NORTH  |  BILLINGS MT 59101

406.259.6387 – ALIGNEDMEDICALSOLUTIONS.COM

 On April 27, 2026, Aligned Medical Solutions, Billings, MT, initiated a nationwide recall of one convenience kit that contains the Huons Pharmacy Lidocaine HCL 1% 5mL SDV. 

Associated kit is AMS16835 Fluids Kit – RX. 

Product(s) can be identified by the pack label inside the sterile barrier of the convenience kit. Product labels can be viewed by clicking here.

Recalled products include:

These convenience kits contain the recalled Huons Pharmacy Lidocaine HCL 1% 5mL SDV, Brand Name – Pharmacy Lidocaine Hydrochloride Injection USP, part #73293-0001-2. Huons Co. Ltd. is voluntarily recalling all lots of NDC #73293-0001-2, Lidocaine Hydrochloride Injection, USP. Huons is initiating this recall due to quality issues identified during a recent FDA manufacturing facility inspection. They began shipping this product on or about April 1, 2023. Use or consumption of this product may include a potential health hazard. 

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=219569

 Aligned Medical Solutions distributed these kits nationwide. Recalled product[s] were manufactured from 12/26/2025 to 4/5/2026 and distributed from 12/26/2025 to 4/6/2026. 

No injuries have been reported to Aligned Medical Solutions to date. 

Customers who have affected product[s] should immediately review their inventory and quarantine all affected kits. Contact the Quality Department for further instructions on labeling and replacement product if needed at 406.259.6387. Customers will be provided with a label that is to be placed on the packs containing the recalled Huons Lidocaine Hydrochloride Injection. Aligned Medical Solutions is notifying its distributors by email and will arrange for labels to identify the packs with the recalled syringe. At the time the kit is opened for use any Lidocaine Hydrochloride Injection should be removed and either returned or destroyed. 

The recalled Huons Lidocaine Hydrochloride Injection should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. All other components in the kit are not affected by this recall. Aligned Medical Solutions will also arrange for replacement Lidocaine if requested. 

Customers with questions may contact the company via telephone at 1.800.360.9428 between the hours of 8:00 am and 5:00 pm [Mountain Time]. Customers may also contact the company via e-mail at fieldcorrectiveaction@alignedmedicalsolutions.com. 

Please contact Michael Cassou at mcassou@alignedmedical.com for replacement product or via telephone 406.259.6387 between the hours of 8:00 am and 5:00 pm [Mountain Time].

Aligned Medical Solutions has notified the FDA of this action. 

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA’s MedWatch Adverse Event Reporting via: 

Complete and submit the report Online: www.fda.gov/medwatch/report.htm 

Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1.800.332.1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1.800.FDA.0178. 

PRESS RELEASE

Windstone Medical Packaging dba Aligned Medical Solutions Issues Nationwide Recall of Namic Angiographic Rotating Adapter 10CC Control Syringe placed into the following kits: AMS6908E Angio Pack and AMS6908F Angio Pack

 

APRIL 13, 2026

1602 4TH AVENUE NORTH  |  BILLINGS MT 59101

406.259.6387 – ALIGNEDMEDICALSOLUTIONS.COM

On April 02, 2026, Aligned Medical Solutions, Billings, MT, initiated a nationwide recall of two convenience kits that contain the Medline Namic Angiographic Rotating Adapter [RA] 10CC Syringe.

Associated kits are AMS6908E Angio Pack and AMS6908F Angio Pack.

Product[s] can be identified by the pack label inside the sterile barrier of the convenience kit. Product labels can be viewed at the following link by clicking here.

Recalled products include:

These convenience kits contain the recalled Medline Namic Angiographic [RA] 10CC Syringe, Brand Name—Medline Medical Industries part #80085007. This product has been recalled because Medline has identified through post-market surveillance, risk of the syringe rotating adapter unwinding during use, which may result in a loose connection and/or full disconnection between the syringes and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, and infection, as well as introduction of air into the line, which may result in air embolism—a condition that carries a serious injury or death. All units have the potential to exhibit this failure mode. There have been three instances reported to Medline of inadvertent air injection into the patient, and one instance of exposure to biohazard, none of these instances have resulted in death.

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=219145

Aligned Medical Solutions distributed these kits nationwide. Recalled product[s] were manufactured from 10/18/2024 to 11/13/2025 and distributed from 10/28/24 to 11/28/2025

No injuries have been reported to Aligned Medical Solutions to date.

Customers who have affected product[s] should immediately review their inventory and quarantine all affected kits. Contact the Quality Department for further instructions on labeling and replacement product if needed at 406.259.6387. Customers will be provided with a label that is to be placed on the packs containing the recalled Medline Namic Angiogrphic [RA] 10CC Syringe. Aligned Medical Solutions is notifying its distributors by email and will arrange for labels to identify the packs with the recalled syringe. At the time the kit is opened for use any MEDLINE 10cc Syringe should be removed and destroyed.

The recalled MEDLINE 10cc Control Syringe should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. All other components in the kit are not affected by this recall.  Aligned Medical Solutions will also arrange for replacement syringes if requested.

Customers with questions may contact the company via telephone at 1.800.360.9428 between the hours of 8:00 am and 5:00 pm [Mountain Time]. Customers may also contact the company via email at fieldcorrectiveaction@alignedmedicalsolutions.com.

Please contact Vicki Davis at vdavis@alignedmedical.com for replacement product or via telephone 406.259.6387 between the hours of 8:00 am and 5:00 pm [Mountain Time].

Aligned Medical Solutions has notified the FDA of this action.

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA’s MedWatch Adverse Event Reporting via:

Complete and submit the report online: www.fda.gov/medwatch/report.htm

Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1.800.332.1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1.800.FDA.0178.

PRESS RELEASE

Windstone Medical Packaging dba Aligned Medical Solutions Issues Nationwide Recall of Nurse Assist Sterile Saline placed into the following kits: AMS10833 Local Lower Extremity Pack I, AMS12947 In House Ocular Pack, AMS12947A In House Ocular Pack and AMS13043 Closure Kit 

JANUARY 22, 2024 

1602 4TH AVENUE NORTH  |  BILLINGS MT 59101 

406.259.6387 – ALIGNEDMEDICALSOLUTIONS.COM

On December 22, 2023, Aligned Medical Solutions, Billings, MT, initiated a nationwide recall of Nurse Assist Sterile Saline which is piggybacked [attached to the outside of the kit in a separate pouch] to the following kits: AMS10833 Local Lower Extremity Pack I, AMS12947 In House Ocular Pack, AMS12947A In House Ocular Pack and AMS13043 Closure Kit. 

Product[s] can be identified by the pack label inside the sterile barrier of the convenience kit. Product labels can be viewed at the following link by clicking here. 

Recalled products include: 

These convenience kits contain the recalled Nurse Assist Sterile Saline, Brand Name – Medline Medical part #DYND40540 110ML Sterile Saline Cup or Nurse Assist Sterile Saline Brand Name – McKesson Medical part #560284 100ML Sterile Saline Bottle, which has been recalled because the Nurse Assist product cannot be verified as having the required sterility assurance level [SAL] of 10-6. See Nurse Assist Recall of 0.9% Sodium Chloride Irrigation USP and Sterile Water for Irrigation 

https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts/nurse-assist-llc-issues-recall-09-sodium-chloride-irrigation-usp-and-sterile-water-irrigation-usp 

Water-based medical products that are nonsterile and potentially contaminated with bacteria could cause serious or life-threatening infections. 

Patients who are elderly, critically ill, have weak immune systems [including newborn infants, pregnant women, and cancer patients] or have chronic diseases are particularly at risk of infection. However, other patients could also develop infections after they are exposed to contaminated water-based medical products. 

Aligned Medical Solutions distributed these kits nationwide. Recalled product[s] were manufactured from 06/02/2022 to 11/20/23 and distributed from 06/08/2022 to 11/27/2023. 

No injuries have been reported to date. 

Customers who have affected product[s] should immediately review their inventory and quarantine all affected kits. Contact the Quality Department for further instructions on labeling and replacement product if needed at 406.259.6387. Customers will be provided with a label that is to be placed on the packs containing the recalled Nurse Assist Sterile Saline. Aligned Medical Solutions is notifying its distributors by email and will arrange for labels to identify the packs with the recalled sterile saline. At the time the kit is opened for use, Nurse Assist sterile saline should be identified, quarantined, and disposed of pursuant to the medical waste policies in effect in your institution. All other components in the kit are not affected by this recall. Aligned Medical Solutions will also arrange for replacement sterile saline to be sent for all recalled product[s]. 

Customers with questions may contact the company via telephone at 1.800.637.7056 between the hours of 8:00 am and 5:00 pm [Mountain Time]. Customers may also contact the company via email at fieldcorrectiveaction@alignedmedicalsolutions.com. 

Please contact your McKesson Sales Rep or Vicki Davis at vdavis@alignedmedical.com | 406.259.6387 for replacement product. 

Aligned Medical Solutions has notified the FDA of this action. 

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA’s MedWatch Adverse Event Reporting via: 

Complete and submit the report online: www.fda.gov/medwatch/report.htm 

Regular mail or fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1.800.332.1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1.800.FDA.0178.

error: Content is protected !!